Issue 67
T. Diburov et alii, Frattura ed Integrità Strutturale, 67 (2024) 259-279; DOI: 10.3221/IGF-ESIS.67.19
96 MPa
(c) Figure 5: The distributions of equivalent stresses (a, c) and strains (b) at the installation of the Zygan implant in the upper left position; (a) general view; (c) stresses in bone tissue. Fig. 6 presents distributions of equivalent stresses and strains at the installation of the Oncology implant in the lower left position. As in all the above-described cases, the maximum eq value of 183 MPa was observed at the “implant–zygomatic bone” interface, which was lower by ~20 MPa than that for the right position according to Fig. 4. A more detailed view of the eq distribution showed that their values did not exceed 128 MPa in the zygomatic bone, which was higher by ~30 MPa than that for the right position. According to Fig. 6, b, the eq maximum values were within 1.4% in bone tissue (in contrast to eq ~1.3% for the Oncology implant shown in Fig. 4, b). The maximum values of equivalent stresses and strains, determined in four performed calculations, are summarized in Tab. 3.
Maximum equivalent stresses and strains
No.
Type and position of implant
in structure, MPa
in zygomatic bone, MPa
in zygomatic bone, %
1
Zygan, the upper left position
263
118
1.6
2
Oncology, the lower right position
201
92
1.3
3
Zygan, the upper left position
244
90
1.1
4
Oncology, the lower left position
183
128
1.4
Table 3: The SSS parameters for the installed single zygomatic implants.
0.014
183 MPa
(a)
(b)
268
Made with FlippingBook Learn more on our blog